APD is not performed when MS1 scan is placed via the IAPI #34
Comments
This was reported previously as well. I happy to say it's fixed in the latest Tune version 4.0 for tribrids. I will try to get some clarity on the situation for Exploris. |
I was not sure the two things were related, not much info were provided in the other open issue :) |
I had a bit of time so I decided to upgrade the Tune and try myself. |
Hi @jesse-canterbury, is there any update on this for the Exploris? Thanks 🤞 |
No updates for the moment, but a resolution will likely have to wait for a new version of Exploris control software. I'll try to get more information. |
@pisistrato What is an APD scan? |
Advanced Peak Determination. Well, no monoisotopes determination is
performed at all on MS1 scans.
…On Thu, Jul 13, 2023, 18:35 Dmitry Avtonomov ***@***.***> wrote:
@pisistrato <https://github.com/pisistrato> What is an APD scan?
—
Reply to this email directly, view it on GitHub
<#34 (comment)>,
or unsubscribe
<https://github.com/notifications/unsubscribe-auth/AJC6DX4IZBRREN25LXSXUVDXQAPVNANCNFSM6AAAAAAUNNOXGM>
.
You are receiving this because you were mentioned.Message ID:
***@***.***>
|
@pisistrato Thank you! |
I have noticed that if the MS1 scan in placed via the IAPI, the returning spectrum is only centroided, the ICentroid.IsMonoisotopic flag is never TRUE.
I was wondering if this is intended and whether it would be possible to apply APD also on custom scans in the future.
Tested on both Eclipse and Exploris, running the latest Tune versions available as of today.
The text was updated successfully, but these errors were encountered: