DrawingRecorder

  1. Start the recorder.
  2. Properly set the correct project path.
  3. Fit the paper onto the tablet, and ensure it doesn't move.
  4. Calibrate:
    • Enter an Operator, Tablet ID, Pen ID and a Drawing ID (when using a barcode gun, zap the tablet, pen and the paper).
    • Move the pen over the red point indicated in the screen.
    • Hold the pen straight. When the point fills yellow, press ENTER.
    • Repeat for all the points, until all points are green and the cursor appears as a crossbar.
    • Review that the pen matches the screen position/drawing by moving it over intersection points.
    • If it doesn't match properly, press TAB to reset.
    • When satisfied, press ENTER.
  5. Press 'New Recording'.
  6. Enter a valid Operator and Patient ID (when using a barcode gun, zap the AID).
  7. Let the patient hold the pen straight and without touching the surface. The patient must hold the other hand on the lap (to avoid moving the table).
  8. Let the patient draw a spiral within the spiral (in the space between the lines), starting from the center and outward.
  9. Review that the drawing has no recording issues.
  10. If the drawing has any problems, press TAB and repeat from step 7.
  11. When satisfied, press ENTER.
  12. Set the patient details (left/right handedness, hand used for drawing, patient/case) and eventually comments when needed.
  13. Save the drawing by pressing "Next hand", which will automatically cycle between hands until all the required spirals have been recorded.
  14. Double-check that the paper still doesn't move. If the paper moves, repeat the calibration procedure from step 3.

Multiple strokes:

There are some holes in the drawing (the pen was lifted while drawing), which might result in a recording which is more difficult to analyze. Consider repeating the recording.

Short recording:

The recording was either very quick, or contained very few events. Check for potential problems.

Old calibration data:

The last calibration was performed more than 8 hours ago and/or more than 50 drawings ago. Ensure that the paper is still properly fitted.
  • Do not run concurrent applications while recording, as doing so may introduce jitter in the recording!
  • Disable any feature of the tablet/pen: patients should not worry about pressing buttons accidentally on the tablet, and doing so should not interrupt the recording (this is described in the installation instructions).
  • Do not change any pressure setting ever in the tablet/Wacom preferences. Follow the recommended configuration instructions precisely.
  • If the sheet of paper moves and/or is wavy, replace it and recalibrate.
  • When calibrating, hold the pen straight, then review the intersection points.
  • The patient should be concentrated on the drawing, not on the screen. Turn away the screen from the patient.
  • Review the drawing when the patient terminates. The drawing should consist of a single line without holes (watch out for unexpected jumps of the cursor, etc). If you see a problem with the drawing, reset with TAB and restart.

Codes that can be used for testing:

  • AID: "0"
  • Tablet ID: "T0"
  • Pen ID: "S0"
  • Drawing ID: "DSPR1"

To see how codes should be constructed, see Coding of IDs.

  • Windows 7/Wacom Intuos: During calibration, when pressing the pen onto the tablet, the point never fills red:

    The tablet might not be working correctly, or might not yet be ready. It seems that the Wacom drivers require up to 60 seconds after login to work. You can verify this problem by going to the control panel and clicking on the Wacom preferences link just after logging in (an error message will appear mentioning that there are no Wacom tablets connected).

    Wait at least 60 seconds after login before starting the Recorder application. Quit and restart the Recorder if necessary.

    If the Recorder still fails to recognize any pressure on the tablet, check in the Wacom preferences that the pen is working correctly.

1.9:

  • DrawingRecorder file format 1.6.

1.8:

  • DrawingRecorder file format 1.5.
  • Added TabletTester to debug tablet-related issues.
  • DrawingRecorder now also includes lower-level packet timestamps/serials on Windows which are better suited for analysis.

1.7:

  • DrawingRecorder file format 1.4.
  • Fixed file encoding issues when saving unicode data.

1.6:

  • DrawingRecorder file format 1.4.
  • DrawingRecorder/StylusProfiler now output JSON-encoded files using the same data structure. Saving/loading times are drastically improved.
  • Improved time resolution of the event stream under Windows.

1.5:

  • DrawingRecorder file format 1.4.
  • drwstats/drwset now produce/parse the same data format and field names.
  • profmap has been added in order to visualize the stylus profile correction over the study time span.
  • surgen can read drwstats output files to generate usage reports that can be input to profmap.
  • DrawingRecorder shows the hand being recorded in the main header after the first patient details have been entered.
  • DrawingRecorder can now require a comment to be given when patient details are being changed by the operator in the same recording session.

1.4:

  • DrawingRecorder file format 1.3.
  • DrawingVisualizer/drwrenderer correctly draw spirals resulting from partial recordings (as generated by pressing RESET mid-flight).
  • DrawingVisualizer samples the drawing speed only once, on demand.
  • DrawingVisualizer shows stroke order/lifting points by default.
  • DrawingVisualizer includes contrast/bias controls for speed and pressure.
  • DrawingVisualizer supports fast-loading.
  • StylusProfiler includes a "tare" field for improved calibration workflow.
  • DrawingRecorder enforces a <5deg tilt while calibrating.
  • DrawingRecorder blinks "RECORDING" when the event stream starts.
  • DrawingRecorder shows the current spiral count/total after each recording.
  • DrawingRecorder requires a "Project file" (config.yaml) to be present in the project path. This file defines the recording settings and patient types, which were previously fixed.
  • DrawingRecorder saves the spirals in a YYYYMM sub-directory of the project path, to avoid directory performance issues on large studies.
  • drwset/profset have been added to upgrade, update or dump attributes within existing drawings/profiles.
  • drwconvert has been removed (superseded by drwset).

1.3:

  • DrawingRecorder file format 1.2.
  • Several tools for data inspection have been added (drwstats, drwrenderer and drwstackrenderer).
  • Tools for data inspection and DrawingVisualizer can now use 'dump' files to speed-up loading/saving time. drwconvert can be used to convert between YaML/dump formats.
  • A simple tool to generate and check IDs with a Verhoeff check digit (patient/table/stylus ID) has been added (genverhoeff).
  • In DrawingVisualizer, the speed is now sampled to give more accurate results.
  • An exception caused by an aborted calibration has been fixed.
  • During calibration, the operator and stylus id are now being requested.
  • All recorded trials (caused by pressing RESET while recording) are now saved.
  • DrawingVisualizer can show recorded trials, when present.
  • Default extension for recordings has been changed to rec.yaml.gz, and a new type record has been added.
  • The prompt dialog at the end of a recording has been extensively revised:
    • The operator id is now also included.
    • Patient handedness, drawing hand and blood drawn status are now mandatory (the user needs to check the appropriate option in all cases)
    • Quality of the preview has been improved.
    • A new option "Next hand" has been added to preserve the patient data and automatically create a recording for the other hand.
    • "Hand cycling" (first hand, second hand, first hand ...) is automatically performed, with 3 cycles being hard-coded, for a total of 6 spirals being requested per-patient.
  • A new tool StylusProfiler has been added:
    • Allows to profile the individual pressure response of each stylus.
    • Performs a simple 3rd degree polynomial fit of the samples.
    • A new file format prof.yaml.gz has been designed for the purpose.

1.2:

  • DrawingRecorder file format 1.1.
  • An exception caused by empty recordings was fixed.
  • Internationalization of the Recorder/Visualizer interface.
  • Add a new checkbox "Blood drawn on drawing arm" after finishing the recording and in the recorded data to reflect our new workflow.
  • An image of the spiral is now shown after performing a recording.
  • The name/id of the operator is now requested for each recording.

1.1:

  • DrawingRecorder file format 1.1.
  • Locale issues under Windows were fixed (notably, DrawingRecorder would refuse to save a recording if the comment contained any accented letter).
  • DrawingRecorder had a glitch that would sometimes cause a failure to start recording (requiring the user to release/press the pen again).
  • Tablet enter/leave events are now also recorded, which improves "trace" tracking as "jumps" are now absent.
  • Improved performance for high-throughput tablets (such as Intuos5).
  • Tilt information is now recorded, both raw and corrected.
  • Added the "DSPR2" drawing ID with the same spiral as DSPR1, but larger sheet for the Intuos5 tablet.
  • 1.0-1.7: Timestamp issues: due to internal limitations, the "stamp" value of the event stream has significant jitter as it's originating from the recorder's event loop itself. Until DR 1.6 the effective resolution on Windows platforms was also lower than 10ms. DR 1.6 introduced higher-quality timestamps on Windows, and DR 1.8 also added "dev_stamp" and "dev_serial" which are suitable for accurate time analysis.
  • 1.0: Tablet enter/leave events not properly tracked: proximity events are still missing from the event stream, meaning that holes in the "trace" require post-processing to be detected, and doing so it not easy due to the quantization of event timestamps. This has been fixed in DrawingRecorder 1.1, but must be kept in mind for files produced by older releases.

As an administrator, install in order:

  • Python 2.7 (python-2.7.3.msi)
  • Python modules (install-modules.bat)
  • PyQt4 (PyQt-Py2.7-x86-gpl-4.9.4-1.exe)
  • PyYAML (PyYAML-3.10.win32-py2.7.exe)
  • NumPy (numpy-MKL-1.8.1.win32-py2.7.exe)
  • SciPy (scipy-0.14.0c1.win32-py2.7.exe)
  • pyqtgraph 0.9.8 (pyqtgraph-0.9.8.win32.exe)
    • Expand pyqtgraph-0.9.8-patches.zip into Python's site-packages path.
    • Run pyqtgraph-recompile.bat.
  • Recompile DrawingRecorder (recompile.bat)

You'll need to use "Run as Administrator" (also for executing a prompt with CMD.EXE) in order to make a system-wide installation.

Customize Windows 7 as follows:

  • Control panel:
    • Pen & touch:
      • Pen options:
        • Disable press & hold
        • Disable visual feedback when touching screen
      • Flicks:
        • Disable flicks
    • Tablet PC settings:
      • Other:
        • Set left/right
        • Input panel settings:
          • Disable "For tablet pen input, show icon next to the text box"
          • Disable "Use the Input Panel tab"

Customize Windows 10 as follows:

  • Control panel:
    • Pen & touch:
      • Pen options:
        • Disable press & hold
        • Disable visual feedback when touching screen
      • Flicks:
        • Disable flicks
    • Tablet PC settings:
      • Other:
        • Set left/right
        • Input panel settings:
          • Disable "For tablet pen input, show icon next to the text box"
          • Disable "Use the Input Panel tab"
    • Settings:
      • Pen & Windows Ink:
        • Disable "Show visual effects"
        • Disable "Show cursor"
        • Enable "Ignore touch input whem I'm using my pen"
        • Disable all "Pen shortcuts"

After performing the common installation/customization procedure, proceed by installing in order:

  • Wacom drivers (cons525-5a_int.exe)

Then customize the tablet preferences:

  • Control panel:

    • Bamboo Preferences:

      • Tablet:
        • Set orientation
        • Disable all "Express Keys"
      • Pen:
        • Disable "Pan/scroll"
        • Mapping:
          • In a single-monitor setup, leave the default.
          • In a dual-monitor setup, set the pen to use the whole area of the screen used for display.
      • Touch options:
        • Disable touch input

After performing the common installation/customization procedure, proceed by installing in order:

  • Wacom drivers (WacomTablet_634-3.exe)

After installing/rebooting, please move the pen over the tablet at least once so that the Wacom driver shows it into the preferences.

Customize the tablet preferences as follows:

  • Control panel:
    • Wacom Tablet Properties:
      • Options:
        • Disable "Pressure compatibility" (important!)
      • Tablet/Functions/All:
        • Express keys:
          • Disable all "Express Keys"
          • Disable "Show Express View"
        • Touch ring:
          • Disable all corners
          • Disable "Show touch ring setting"
      • Tablet/Touch/All:
        • Touch options:
          • Disable touch input
      • Tablet/Grip pen/All:
        • Pen:
          • Disable buttons (double/right click)
        • Eraser:
          • Disable eraser
        • Mapping:
          • Set orientation (usually "ExpressKeys Left")
          • Screen area:
            • In a single-monitor setup, leave the default.
            • In a dual-monitor setup, set the pen to use the whole area of the screen used for display.

After performing the common installation/customization procedure, proceed by installing in order:

  • Wacom drivers (WacomTablet_634-3.exe)

After installing/rebooting, please move the pen over the tablet at least once so that the Wacom driver shows it into the preferences.

Update the firmware to the latest release and don't upgrade anymore during the study.

Disable the touch input using the hardware switch on the left.

Customize the tablet preferences as follows:

  • Wacom Desktop Center:
    • Disable AutoStart
  • Control panel:
    • Wacom Tablet Properties:
      • Options:
        • Disable "Pressure compatibility" (important!)
        • "Disable All Tablet buttons"
        • Disable "Show wireless tablet battery status"
      • Tablet/Functions/All:
        • Express keys:
          • Disable all "Express Keys"
          • Disable "Show Express View"
        • Touch ring:
          • Disable all corners
          • Disable "Show touch ring setting"
      • Tablet/Touch/All:
        • Touch options:
          • Disable touch input
      • Tablet/Grip pen/All:
        • Pen:
          • Disable buttons (double/right click)
        • Eraser:
          • Disable eraser
        • Mapping:
          • Set orientation (usually "ExpressKeys Left")
          • Screen area:
            • In a single-monitor setup, leave the default.
            • In a dual-monitor setup, set the pen to use the whole area of the screen used for display.
          • Disable "Use Windows Ink"

The analysis modules will work with just the reported pressure, but for comparable measures the real applied weight is required. StylusProfiler is a simple tool that allows to build a response curve of the stylus.

Important

Never change the configuration settings for the stylus/tablet during the study; especially the settings for the pressure feel, as it would obviously make values incomparable.

Brand-new styluses will quickly drift between each single measurement, making it impossible to get a reliable profile. The styluses need to be used for at least ~20 recordings in order to get stable measurements.

It's recommended to perform a response profile as often as possible in order to capture the response variability, especially during the first week of usage. The analysis modules can load multiple profiles at different time points and interpolate to get a reasonable estimate of the real weight.

Start StylusProfiler and insert the basic parameters (operator, tablet/stylus ID).

The data can be inserted either manually (by setting both "Pressure" and "Weight") or by inserting "Pressures" as directly measured by the tablet by putting the stylus on top of the tablet. When the stylus is on the tablet the mouse cannot be used (use the key accelerators or TAB to move between fields).

Insert a (0,0) data-point manually first.

Put the stylus on a scale and measure/enter its weight. Put the calibrator base on the tablet, then put the stylus in it by dropping it 2cm above its resting position (to reduce nib friction). The "Pressure" field is automatically populated. Press "Add" once the pressure is stable. This first measurement constitutes the lightest measurable pressure.

Put the stylus in the reversed pusher plate, then measure both on a scale. Enable the "Tare" field by pressing the "T" button and enter the base weight which will also serve for further measurements. Put the stylus/plate onto the base. Enter "0" as a weight and add the data-point.

Use a liquid in a light plastic container (a regular plastic cup is fine) to measure the response at 25g intervals in 2 ramps:

  • Start by measuring from the base weight in 50g intervals until the upper limit is reached (the reported pressure is the closest < 1).
  • Hold the base while applying heavy weights on the pusher plate.
  • Restart from the base + 25g and measure the remaining data-points at 50g intervals.

This protocol will reduce cumulative errors introduced by the response drift in the upper range. When measuring the reported pressure, be sure to release the weight on top the stylus quickly (it's easily done by dropping the weight from 1-2cm above the pusher plate).

An easy way to check each measurement for potential drift is to lift/drop the pusher plate before applying the weight, and check that the previous reference value is still matching.

After all the 25g intervals are taken, to measure weights lower than the pusher plate, use the tubes 2, 1+4, 1+2+5 in sequence by placing them onto the stylus and measuring pen+tubes together first on a scale, then onto the tablet. Be sure to disable the "Tare" (these are direct measurements). Each tube combination increases by approximately 10g.

The file formats are stored in self-descriptive GZip-compressed JSON. Versions of DrawingRecorder <= 1.5 used the same structure but YaML encoding. GZip is used both to conserve space and for check-summing purposes.

To speed-up loading for repeated processing, drwset can be used to convert an existing file into a "dump" object that loads faster. It's important to note though that such dumps must not be used for distribution and are not compatible across different versions.

drwset/profset can also export raw, unprocessed values into tabular text files for other (usually debugging) purposes.

The project configuration file is required to be present in the project directory, where the recorded spirals are also stored. It must be named as config.yaml and includes recording settings, patient types and study details which are used for all subsequent recordings.

A sample file (with explanatory comments) is included in recordings/config.yaml which details all the required keys. Copy/customize this sample file as required.

Keys related to drawing/calibration (all keys are mandatory):

  • drawing/points: contains a list of coordinate pairs (from now on: points) in "normalized drawing space" that represent the drawing as overlaid on the paper (the spiral itself).
  • drawing/cpoints: contains a list of points in "normalized drawing space". Each point is used as a calibration target, and is mapped to a different coordinate space in calibration/cpoints at the same list index.
  • calibration/cpoints: contains a list of points, with each point being a calibration target for drawing/cpoints but in the same coordinate space as recording/events/cdraw.
  • recording/events: each event has at least two point pairs: cdraw and ctrans:
    • cdraw: points in the same coordinate space as calibration/cpoints.
    • ctrans: points in recorder's internal viewing space.
  • recording/rect_drawing: contains the screen quadrilateral in the same coordinate space as calibration/cpoints.
  • recording/rect_trans: contains the screen quadrilateral in the same coordinate space as recording/events/ctrans.
  • recording/rect_size: the size of the screen (in pixels) during the recording.

Ancillary data (all keys are mandatory):

  • format: file format version (1.* describes this format)
  • version: application version
  • aid: patient AID
  • drawing/id: drawing ID
  • drawing/str: drawing description (redundant for human readability)
  • calibration/tablet_id: tablet ID used for calibration
  • calibration/stamp: timestamp of the last calibration
  • calibration_age: number of drawings since the last calibration
  • recording/session_start: timestamp of the start of the session (when the recording window is initially shown)
  • recording/retries: number of attempts required for a correct recording
  • recording/strokes: number of strokes in the recording (redundant for human readability)
  • pat_type: patient type
  • pat_handedness: patient handedness
  • pat_hand: patient hand
  • comments: free text comment for the recording

Chunks introduced with format 1.1:

  • recording/events:
    • tdraw (optional): uncorrected x/y tilt information expressed in +/- 0-60 degrees for each axis.
    • ttrans (optional): rotation-adjusted x/y tilt information.
  • extra_data:
    • blood_drawn (optional): reflects the new "Blood drawn on drawing arm" introduced in DrawingRecorder 1.2.
    • operator (optional): the name of the operator assisting during the recording (introduced in DrawingRecorder 1.2, moved in 1.4).

Chunks introduced with format 1.2:

  • type: to distinguish file types (recording/profiles), type has been added, and needs to be rec when present.
  • calibration/stylus_id: stylus ID (introduced in DrawingRecorder 1.3)
  • calibration/operator: operator performing the calibration (introduced in DrawingRecorder 1.3)
  • recording/retries_events: An ''array'' of events with the same data and format as recording/events, one for each trial during the recording. recording/retries is just the length of this array + 1 (for backward compatibility).
  • pat_hand_cnt: number of hands the patient is able to draw with.
  • cycle: cycle number in a single recording session.

Chunks introduced with format 1.3:

  • ts_created: drawing creation timestamp

  • ts_updated: drawing update (last change) timestamp

  • operator: the name of the operator assisting during the recording (moved from extra_data/operator in DrawingRecorder 1.4).

  • extra_data:

    These fields were introduced in DrawingRecorder 1.4 and moved in 1.5:

    • orig_format: original file format version before the file has been re-saved. This field is created by drwset when a file has been upgraded from an older format, and is never overwritten.
    • orig_version: original application version version before the file has been re-saved. This field is created by drwset when a file has been upgraded from an older format, and is never overwritten.
    • orig_pat_type: original patient type before being modified by drwset. This field is created by drwset when the patient type has been manually reset and is never overwritten.

Chunks introduced with format 1.4:

  • tz: timezone offset in seconds from UTC

  • extra_data/orig:

    Dictionary of initial values which have been changed when writing/updating an existing file. Existing values are never replaced. Keys include:

    • format moved from extra_data/orig_format in DrawingRecorder 1.5.
    • version moved from extra_data/orig_version in DrawingRecorder 1.5.
    • Any field name from the output of drwstats.

Chunks introduced with format 1.5:

  • recording/events:
    • dev_stamp (optional): accurate event timestamp (in MS)
    • dev_serial (optional): event serial number

Chunks introduced with format 1.6:

  • drawing/area_dim: Physical dimensions of the drawing area (in mm).

Keys related to the profile (all keys are mandatory):

  • data: list of data points, where each point contains:
    • press: stylus pressure reported
    • weight: applied weight
  • fit: 3rd degree polynomial fit of the response curve. Can be null.

Ancillary data (all keys are mandatory):

  • format: file format version (1.* describes this format)
  • type: "prof".
  • version: application version
  • tablet_id: tablet ID used for calibration
  • operator: the name of the operator performing the calibration
  • stylus_id: stylus ID currently being profiled
  • ts_created: profile creation timestamp
  • ts_updated: profile update (last change) timestamp
  • extra_data: provisional dictionary for arbitrary data.

Chunks introduced with format 1.4:

  • tz: timezone offset in seconds from UTC

Several coordinate types and transformations are stored in the file itself.

Coordinates that come directly from the tablet are mapped onto the screen (with range 0x0 to screen's WxH). Since the tablet has a higher resolution than that of the screen, the resulting coordinates are floating point.

When the user draws on the tablet during the calibration, the coordinates are re-mapped so that the center of the tablet matches center of the drawing with an unit-less scale and a square aspect ratio. This is the "drawing space" (as stored in calibration/cpoints and recording/events/cdraw).

The "normalized drawing space" uses information from the calibration points to map the drawing to the unit length and direction using an affine transform. By using such mapping it's possible to reconstruct the original drawing unit.

AID codes in the spirography software must be an all-numeric Verhoeff code. "0" can be used here for testing purposes (which is still valid Verhoeff).

A tablet ID follows the pattern Txxxyyyz where:

  • T: mandatory
  • xxx: study code
  • yyy: incremental code
  • z: Verhoeff check digit

"T0" can be used for testing purposes.

A pen/stylus ID follows the pattern Sxxxyyyz where:

  • S: mandatory
  • xxx: study code
  • yyy: incremental code
  • z: Verhoeff check digit

"S0" can be used for testing purposes.

All drawing IDs currently begin with D have the structure Dxxxy, where:

  • D: mandatory
  • xxx: drawing type
  • y: drawing number

Drawing IDs do not require a Verhoeff check digit, as the list of IDs is always know to the recorder module.

The blueprints for the drawings are stored in the "drw/" directory in the source code. Each drawing type is currently handled by a separated drawing module, since the module itself contains the logic for proper calibration.

  • Either fix PyQt4 to supplement device's timestamp to the QTabletEvent class, or use the pyglet's "wintab" module on Windows, which doesn't require re-compiling/patching PyQt.
  • More drawing types (CCW, two spiral module, etc).
  • Multiple drawings in a single session require rethinking a bit the output format (drawing/points needs to be a list of lists) and recording itself (do we want to perform drawing separation ourselves, or not?).
  • Implement a batch analysis module.
  • Record the actual tablet serial/details in the file instead of relying on the user scanning a barcode.
  • At least on Linux with the Wacom driver, the Intuos Pro 5L tablet produces a variable event rate which combines both event filtering (reducing the event rate while tracking), but also having burst rates higher than the 130.320Hz of the previous model (measured up to ~200Hz). This requires rethinking the filtering/processing code.